154BC Series Pressure Transducer

154BC Series Pressure Transducer Description: The 154BC is a low cost unit designed without a header for o-ring mounting and OEM applications where compatibility with corrosive media is required. The titanium option offers greater resistance to corrosive materials and harsh cleaning chemicals, opening the range of its application use to things such […]

154BC Series Pressure Transducer

Description: The 154BC is a low cost unit designed without a header for o-ring mounting and OEM applications where compatibility with corrosive media is required. The titanium option offers greater resistance to corrosive materials and harsh cleaning chemicals, opening the range of its application use to things such as dialysis machines and salt/brackish water measurements. The sensing package utilizes silicone oil to transfer pressure from either a 316L stainless steel or titanium diaphragm to the sensing element. A ceramic substrate is attached to the package that contains laser-trimmed resistors for temperature compensation and offset correction. An additional laser-trimmed resistor is included which can be used to adjust an external differential amplifier and provide span interchangeability to within ±1%. Please refer to the 154N uncompensated and constant voltage datasheets for more information on different features of the 154.

  • O-Ring Mount
  • -20ºC to 85ºC Compensated Temperature
  • 1.0% Interchangeable Span 1. (provided by gain set resistor)
  • Solid State Reliability
  • ±0.3% Pressure Non Linearity
  • Titanium Option

 

PARAMETERS MIN TYP MAX UNITS NOTES
Span 75 100 150 mV 1
Zero Pressure Output -1.0   0 1.0 mV 2
Pressure Non Linearity -0.30 0.30 %Span 3
Pressure Hysteresis -0.20 0.20 %Span
Repeatability ±0.02 %Span
Input Resistance 2.0 3.5 5.8 3
Output Resistance 3.0 6.0
Temperature Error – Span -1.0 1.0 %Span 4
Temperature Error – Offset -1.0 1.0 %Span 4
Thermal Hysteresis – Span -0.25 ±0.05 0.25 %Span 4
Thermal Hysteresis – Offset -0.25 ±0.05 0.25 %Span 4
Long Term Stability – Span ±0.10 %Span/Year
Long Term Stability – Offset ±0.25 %Span/Year
Supply Current 0.5 1.5 2.0 mA 5
Output Load Resistance 5 6
Insulation Resistance (50Vdc) 50 7
Output Noise (10Hz to 1kHz) 1.0 uV p-p
Response Time (10% to 90%) 0.1 ms
Overload Pressure 2X Rated 8
Burst Pressure 3X Rated 9
Compensated Temperature -20 85 °C
Operating Temperature -40 125 ºC 10
Storage Temperature -50 125 ºC 10
Media – Pressure Port Liquids
and Gases compatible with 316L
Stainless Steel or ASTM Grade
2 CP Titanium

Notes
1. For amplified output circuits, 3.012V ±1% interchangeability with gain set resistor. See application schematic.
2. Measured at vacuum for absolute (A), ambient for gage (G).
3. Best fit straight line.
4. Over the compensated temperature range with respect to 25°C.
5. Guarantees output/input ratiometricity.
6. Load resistance to reduce measurement errors due to output loading.
7. Between case and sensing element.
8. 2X or 500psi, whichever is less. The maximum pressure that can be applied without changing the transducer’s performance or accuracy.
9. 3X or 600psi, whichever is less. The maximum pressure that can be applied to a transducer without rupture of either the sensing element
or transducer.
10. Maximum temperature range for product with standard cable and connector is -20°C to 105°C.

No Information Available.

-20ºC to 85ºC Compensated Temperature
Maximum temperature range for product with standard cable and connector is -20°C to 105°C.