1630 Series Pressure Transducer

1630 Series Pressure Transducer Description: The 1630 is a fully piezoresistive silicon pressure sensor for use in intrauterine pressure monitoring. The sensor is designed to be used with automated assembly equipment and can be dropped directly into a customer’s disposable intrauterine housing. The sensor is designed to meet the requirements as […]

1630 Series Pressure Transducer

Description: The 1630 is a fully piezoresistive silicon pressure sensor for use in intrauterine pressure monitoring. The sensor is designed to be used with automated assembly equipment and can be dropped directly into a customer’s disposable intrauterine housing. The sensor is designed to meet the requirements as described in the Association for the Advancement of Medical Instrumentation (AAMI) specification for Blood Pressure Transducers. The pressure sensor consists of a pressure sensing element mounted on a ceramic substrate. Thick film resistors on the ceramic substrate are laser trimmed for compensation and calibration. A plastic cap is attached to the ceramic substrate to provide an easy method of attachment to the customer’s assembly and protection for the sensing element. A dielectric gel is placed over the sensor to provide electrical and fluid isolation. The 1630 pressure sensors are batch manufactured in a 3×8 element array on a ceramic substrate (24 units per substrate). The products are shipped in anti-static shipping containers. They can also be shipped on a tape and reel. Performance characteristics and packaging can be easily tailored on a special order basis to meet the requirements of specific customers.

Features:
• Disposable Blood Pressure Sensor
• AAMI Specifications
• Low Cost Disposable Medical Applications
• Dielectric Gel Barrier
• Fully Tested and Calibrated
• Small Size and Reliable Performance
• Gel Isolation for Liquids
• Operates from 5°C to 45°C
• Compatible for Automated Assembly
• 1% Accuracy for Replacements
• 5.0 uV/V/mmHg Sensitivity
• Customization for OEM Applications

Supply Voltage: 5.0 Vdc
Ambient Temperature: 23°C (unless otherwise specified)
PARAMETERS MIN TYP MAX

UNITS

NOTES

Operating Pressure Range

-50

300

mmHg

Over Pressure

-400

1200

mmHg

Zero Pressure Offset

-20

20

mmHg

Sensitivity

4.9

5.0

5.1

uV/V/mmHg

Linearity and Hysteresis (0 to 300 mmHg)

-2.0

2.0

%Span

1

Input Impedance

1200

3200

Output Impedance

270

330

Supply Voltage

2

5

10

Vdc or Vac rms

Leakage Current (@ 120 Vac rms, 60Hz)

2

uA

Warm-Up Time

5

Seconds

Frequency Response

1200

Hz

Offset Drift

2

mmHg

2

Thermal Span Shift

-0.1

0.1

%/°C

3

Thermal Offset Shift

-0.3

0.3

mmHg/ C

3

Phase Shift (@ 5KHz)

5

Degrees

Sterilization (ETO)

3

Cycles

4

Operating Temperature

10

40

C

Storage Temperature

-25

70

C

Operating Product Life

72

Hours

Shelf Life

3

Years

Humidity (External) 10-90% (non-condensing)
Media Interface Dielectric Gel

No Information Available.

No Information Available.