N3AIA11A – Load Cell – Lumbar Spine

Load Cell, 3-axial
Location: Lumbar Spine

Technical description

The applied force causes compression or strain of the base body. The deformation is measured using strain gauges. The wiring of multiple strain gauges for a full bridge circuit compensates for the temperature influence on the zero signal and the cross-influence from other force and torque application.

Force direction

Fx, Fz, My

Application

H3-50%
H3-95%

Equivalent types

Denton: 1842
FTSS: IF-402

Measurement specification

Resistive
Strain gauges

Options

ID-Module integrated in sensor
Polarity according to SAE J211

Technical specification

Unit

Value

FX

FZ

MY

Measuring range kN

Nm

13.5

 

13.5

 

 

564

Sensitivity 1) µV/V/kN

µV/V/Nm

89

 

89

 

 

1.2

Output signal 1), 2) mV/V

1.2

1.2

0.7

Bridge resistance Ω

350

350

700

Zero signal 1) mV/V ≤ 0.05
Amplitude non-linearity 3) % ≤ 1.0
Hysteresis 3) % ≤ 1.0
Channel cross talk 3) % ≤ 5.0
Supply voltage V 2–15
Ultimate load % 150
Insulation resistance > 100
Temperature range °C -30…+70
Weight (approximate) g 670/7200
(without/with mass block)

All values measured at 10 V sensor supply voltage and at 23 °C.
1) Typical value
2) At nominal load
3) Relative nominal range

N3AIA11A Dimensions

Temperature range °C -30…+70