N4ATA11A – Load Cell – Lower Tibia

Load Cell, 4-axial
Location: Tibia, Lower

Technical description

The applied force causes compression or strain of the base body. The deformation is measured using strain gauges. The wiring of multiple strain gauges for a full bridge circuit compensates for the temperature influence on the zero signal and the cross-influence from other force and torque application.

Force direction

Fx, Fz, Mx, My

Application

H3-5%, H3-50%, H3-95%
FAA
BioSID

Equivalent types

Denton: 3287J
FTSS: IF-825

Measurement specification

Resistive
Strain gauges

Options

ID-Module integrated in sensor
Polarity according to SAE J211

Technical specification

Unit

Value

FX

FZ

MX

MY

Measuring range kN

Nm

11

 

11

 

 

395

 

395

Sensitivity 1) µV/V/kN

µV/V/Nm

182

 

91

 

 

7.6

 

7.6

Output signal 1), 2) mV/V

2.0

1.0

3.0

3.0

Bridge resistance Ω

350

700

350

350

Zero signal 1) mV/V ≤ 0.05
Amplitude non-linearity 3) % ≤ 1.0
Hysteresis 3) % ≤ 1.0
Channel cross talk 3) % ≤ 5.0
Supply voltage V 2–15
Ultimate load % 150
Insulation resistance > 100
Temperature range °C -30…+70
Weight (approximate) g 500

All values measured at 10 V sensor supply voltage and at 23 °C.
1) Typical value
2) At nominal load
3) Relative nominal range

N4ATA11A Dimensions

Temperature range °C -30…+70