N5ATA11A – Load Cell – Lower Tibia

Load Cell, 5-axial
Location: Tibia, Lower

Technical description

The applied force causes compression or strain of the base body. The deformation is measured using strain gauges. The wiring of multiple strain gauges for a full bridge circuit compensates for the temperature influence on the zero signal and the cross-influence from other force and torque application.

Force direction

Fx, Fy, Fz, Mx, My

Application

H3-5%, H3-50%, H3-95%
FAA
BioSID

Equivalent types

Denton: 3644
FTSS: IF-853

Measurement specification

Resistive
Strain gauges

Options

ID-Module integrated in sensor
Polarity according to SAE J211

Technical specification

Unit Value
FX FY FZ MX MY
Measuring range kN

Nm

11

 

11

 

11

 

 

390

 

390

Sensitivity 1) µV/V/kN

µV/V/Nm

180

 

180

 

95

 

 

8.0

 

8.0

Output signal 1), 2) mV/V 2.0 2.0 1.0 3.0 3.0
Bridge resistance Ω 350 350 700 350 350
Zero signal 1) mV/V ≤ 0.05
Amplitude non-linearity 3) % ≤ 1.0
Hysteresis 3) % ≤ 1.0
Channel cross talk 3) % ≤ 5.0
Supply voltage V 2–15
Ultimate load % 150
Insulation resistance > 100
Temperature range °C -30…+70
Weight (approximate) g 500

All values measured at 10 V sensor supply voltage and at 23 °C.
1) Typical value
2) At nominal load
3) Relative nominal range

N5ATA11A Dimensions

Temperature range °C -30…+70